The FDA said the Tandem Mobi insulin pump recall is the most serious type, with 1,248 serious injuries reported by Tandem Diabetes as of October 28, 2026.

Key points

  1. The FDA identified the Tandem Mobi insulin pump recall as the most serious type, with 1,248 serious injuries reported as of October 28, 2026.
  2. The recall covers all Tandem Mobi Insulin Pumps with Interoperable Technology, with different steps required based on the pump serial number.
  3. Owners whose pumps are on the replacement list should request a replacement right away; others may keep using the device while following updated instructions.
  4. The recall, numbered Z-0316-2027, covers 80,938 devices in commerce and remains Open, Classified in the FDA database.

The FDA has placed a recall of Tandem Diabetes's Tandem Mobi insulin pump in the highest severity category the agency uses. The defect can halt insulin delivery even when no blockage is present. The FDA consumer page on this action said the agency has identified it as the most serious type of recall. As of October 28, 2026, Tandem Diabetes had reported no deaths and 1,248 serious injuries to the FDA.

The FDA's device recall database classifies this as a Class 1 action, recall number Z-0316-2027. Tandem Diabetes initiated the recall on September 23, 2026, and the FDA posted the entry on October 30, 2026. A total of 80,938 devices were in commerce at that time.

Key facts

  • The FDA said it has identified this recall as the most serious type the agency assigns to a medical device action.
  • Tandem Diabetes had reported no deaths and 1,248 serious injuries to the FDA as of October 28, 2026.
  • All Tandem Mobi Insulin Pumps with Interoperable Technology are covered, but required steps differ by pump serial number.
  • Tandem Diabetes sent customer letters on September 23, 2026, one for owners whose pumps need replacement and one for all other Tandem Mobi users.
  • Distribution covered the U.S. nationwide and included Canada, and the recall status is Open, Classified.

What the FDA said about the Tandem Mobi insulin pump

The Tandem Mobi insulin pump is designed to deliver insulin under the skin to help manage diabetes. It connects with compatible continuous glucose monitors and automated insulin dosing systems, allowing the device to adjust delivery automatically.

The issue behind this recall is a false-positive occlusion alarm. An occlusion alarm is meant to alert the user that insulin flow is physically blocked. The FDA said Tandem Diabetes stated that certain pumps may generate this alarm even when no actual obstruction is present. Each time the alarm fires, the pump stops delivering insulin until the user notices the alert, acknowledges it and resumes therapy.

Advertisement

The FDA said repeated alarms that go unaddressed can cause interruptions that push blood glucose higher. When levels rise far enough, hyperglycemia can set in. The agency said that in serious cases, hyperglycemia can require medical treatment or hospitalization. That risk is why the FDA placed this recall in its highest severity category.

What it means for you

All Tandem Mobi Insulin Pumps with Interoperable Technology are included in this action. The FDA said each device in the recall falls into one of two groups, with the required steps determined by the pump's serial number, so the specific action depends on which pump a person owns.

Some pumps must be returned and replaced. Others may stay in service as long as users follow updated instructions. How many of the 80,938 devices in commerce fall into the replacement group is not publicly known. The FDA database lists the serial numbers flagged for exchange, but no published total for that group has been released.

The recall's scope is wide. The FDA database entry listed 80,938 devices in commerce, with 77,559 in the U.S. and 3,379 outside the U.S. That coverage includes nationwide U.S. distribution and Canada.

What you can do

Tandem Diabetes sent two sets of letters to customers on September 23, 2026. One went to owners whose pumps require replacement; the other addressed all remaining Tandem Mobi users. The first step for any user is to check their pump's serial number against the replacement list in the Tandem Diabetes customer letter.

If your serial number is on the replacement list, submit a replacement request right away. Use the link in the customer letter or contact Tandem Customer Technical Support to do so. While waiting, respond to every occlusion alarm and work through the User Guide troubleshooting steps. That means looking over the tubing, cartridge and infusion site for any sign of blockage or damage, fixing whatever is found and resuming delivery. Keep a backup method of insulin delivery available, as your healthcare provider directs.

Advertisement

If your serial number is not on the replacement list, the FDA said you may keep using the pump while following updated instructions. Respond to every alarm, work through the troubleshooting process and resume delivery after each one. Enable push notifications in the Tandem Mobi mobile app as well. That ensures alarm alerts reach your phone in addition to the pump's audible tones or vibrations.

U.S. customers with adverse reactions or quality concerns can reach Tandem Diabetes at [email protected]. The FDA's contact details and further guidance appear on the FDA recall consumer page.

What happens next

The FDA database entry for recall Z-0316-2027 listed its status as Open, Classified at the time it was reviewed. That means the recall is active and the agency is still tracking whether corrective actions are working. No completion date has been announced.

The injury total of 1,248 serious injuries reflects what Tandem Diabetes had filed with the FDA as of October 28, 2026. That figure may change as the recall proceeds and additional reports are filed. The FDA has not published a more recent count.

FAQ

How did the FDA describe the seriousness of this recall?

The FDA consumer page said the agency has identified this recall as the most serious type.

How many serious injuries has Tandem Diabetes reported?

The FDA said Tandem Diabetes had reported no deaths and 1,248 serious injuries associated with this issue as of October 28, 2026.

Who is covered by this recall?

All Tandem Mobi Insulin Pumps with Interoperable Technology are included. The FDA said recommendations vary by pump serial number. Tandem Diabetes sent letters on September 23, 2026 explaining the steps for each group of users. Customers can check their serial number against the replacement list in that Tandem Diabetes letter.

What could happen if alarms are ignored?

The FDA said repeated false-positive occlusion alarms that are not addressed can interrupt insulin delivery and push blood glucose higher. The agency said that when hyperglycemia becomes severe, it can require medical treatment or hospitalization. Users experiencing an unusually high number of alarms even after troubleshooting should contact Tandem Diabetes.